Abstract
A piezo based optical scanner for high speed optical coherence tomography (OCT) has been developed. Design of the miniature probe has been optimized to take advantages of the high speed depth scanning that can be achieved with spectral domain OCT (SD-OCT) systems. Three prototype scanning probes were constructed and tested for OCT imaging. Movies and images of tissues are presented which were taken at frame rates ranging from 50~100fps. Data were acquired at 18000 A-lines per second at an effective integration time of 55us.